Cart 0

Philips Respironics Trilogy 100 Portable Ventilator

RM 48,300.00


Philips Respironics Trilogy 100 Portable Ventilator

Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute- care use to provide invasive and non-invasive ventilator support for adult and pediatric patients. A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology.

Specifications

Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)

Weight
  • 5 kg (11 lb)

Controls
Circuit types
  • Active with PAP; Passive

Ventilation types
  • Pressure and Volume

Volume modes
  • AC; SIMV (w/PS); CV; AC-MPV

Pressure modes
  • CPAP; S; S/T; T; PC; PC-SIMV (w/PS); PC-MPV; AVAPS-AE

Pressure
  • 4 to 50 cm H₂O

IPAP
  • 4 to 50 cm H₂O

PEEP/EPAP for active circuits
  • 0 to 25 cm H₂O

PEEP/EPAP for passive circuits
  • 4 to 25 cm H₂O

CPAP for passive circuits
  • 4 to 20 cm H₂O

Pressure support differential
  • 0 to 40 cm H₂O

Alarms
Apnea rate
  • Off; 10 to 60 sec

Circuit disconnect
  • Off; 5 to 60 sec

High tidal volume
  • Off; 50 to 2000 ml

Low tidal volume
  • Off; 40 to 2000 ml

High minute ventilation
  • Off; 1.0 to 99 l/min

Low minute ventilations
  • Off; 0.2 to 99 l/min

High respiratory rate
  • Off; 4 to 80 BPM

Low respiratory rate
  • Off; 4 to 80 BPM

Measured patient parameters
Respiratory rate
  • 0 to 80 BPM

Minute ventilation
  • 0 to 99 l/min

Vte/Vti
  • 0 to 2000 ml

Leak rate
  • 0 to 200 l/min

Peak inspiratory flow
  • 0 to 200 l/min

Peak inspiratory pressure
  • 0 to 99 cm H₂O

Mean airway pressure
  • 0 to 99 cm H₂O

% Patient triggered breaths
  • 0 to 100 %

I:E ratio
  • 9.9:1 to 1:9.9

Controls
Tidal volume
  • 50 to 2000 ml

Breath rate
  • 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes

Inspiratory time
  • 0.3 to 5.0 sec

Rise time
  • 1 to 6

Ramp start pressure
  • 0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP

Ramp length
  • Off; 5 to 45 min

Flex
  • Off; 1 to 3

Flow trigger sensitivity
  • 1 to 9 l/min

Flow cycle
  • 10 to 90 %

Apnea rate
  • 4 to 60 BPM